Checking Sampling Line Losses for Ozone Measurements
On this page
Quick answer Compare ozone measurements with and without a documented inlet arrangement under a controlled, approved method. Retain the exact instrument, tubing, source and conditions with the result. A published comparison can demonstrate the question to investigate, while suitability for another monitor or sampling path requires its own evidence.
Define what the comparison measures
A tubing check can investigate a concentration difference, a changed baseline, a delayed rise or a longer recovery. State which question the evidence answers. A steady concentration comparison leaves the time required to reach that state to be established separately.
The general sampling-path guide covers transport and instrument response. For ozone, preserve the gas-contact materials and component condition as well. A pump delivering gas does not establish that the intended ozone concentration reaches the measuring cell.
Read a real no-tube and connected-tube comparison
2B Technologies Technical Note 012, dated May 16, 2008, reports an actual manufacturer experiment using a calibrated Model 205, serial 613DB. It compared manifold ozone over 0–300 ppbv without inlet tubing and with 50 ft, stated as 15.2 m, of Teflon-lined Tygon tubing. The stated tubing dimensions were 1/4-inch outside and 1/8-inch inside diameter, with ten measurements at each concentration.
| Reported observation | No inlet tube | Specified 50-ft inlet tube |
|---|---|---|
| Detection-cell pressure | 811.8 mbar | 752.8 mbar |
| Measured volumetric flow | 2.67 L/min | 2.37 L/min |
| Concentration comparison | x, measured ozone without the tube | y, measured ozone with the tube, reported fit y = 0.9970x − 1.9945 ppbv |
The authors regarded the slope-related loss as within measurement error and identified an approximately −2 ppbv offset. These are historical observations on that setup and new tubing. The note also warns that contamination and condensation can change a long inlet's performance.
The equation is a reported comparison summary. It supplies no individual raw readings and gives no rise-time or recovery-time trace. Preserve that distinction when deciding whether this study answers the question posed for another installation.
Review the limits before transferring the result
Identify the exact offered instrument and measurement principle. Record the proposed tube material, dimensions, age and condition, plus every fitting, filter, trap and adapter. Ask which document establishes the permitted complete arrangement for ozone.
The historical trial's material description includes a liner and outer tube. Keep the complete construction in the record. A visually similar flexible tube or a shared trade-name fragment leaves gas-contact identity unresolved.
Obtain the present model's sample and environmental requirements. A later 2B Model 106-L manual, Revision J-3, has its own inlet and ozone-loss checks. Retain such instructions with their actual model instead of combining them with the earlier Model 205 experiment into a new procedure.
Prepare a paired comparison record for qualified review
Have the responsible provider define the gas source, permitted connection, stabilization conditions and acceptance criteria before an actual test. Keep gas generation, application and disposal under the approved procedure. This worksheet organizes the evidence needed by that reviewer.
| Comparison field | Baseline configuration | Connected configuration |
|---|---|---|
| Instrument and reference | Model/serial ____ Verification record ____ | Same instrument or documented comparator ____ |
| Gas and conditions | Source/level/basis ____ Temperature/humidity ____ | Matched conditions ____ Departures ____ |
| Gas-contact path | Required baseline inlet components ____ | Tube/material/length/diameters ____ All accessory IDs ____ |
| Measurement state | Approved stability criterion ____ Individual readings/times ____ | Same criterion ____ Individual readings/times ____ |
| Additional observations | Pressure/flow if required ____ Zero/status ____ | Pressure/flow ____ Zero/status ____ |
| Decision | Criterion/document ____ Reference uncertainty ____ | Difference/uncertainty ____ Outcome ____ Reviewer ____ |
Preserve individual observations and the order of comparison. Record any source variation, changed condition or interrupted run. A summary average alone can hide a changing source or a response that was still developing.
When the intended task also depends on response timing, add a separate approved time-trace comparison. Identify the start event, logging interval, arrival, stabilization and recovery criteria. Leave those outcomes unestablished when the evidence contains only steady measurements.
Reopen the review after a path change
Keep the accepted inlet configuration with its evidence and service history. A new tube, filter, connector or conditioning arrangement creates a changed configuration for review. Record contamination, visible liquid and uncertain component identity through the applicable maintenance procedure.

Use Mcooh's gas detector collection with an enquiry stating ozone range, measurement purpose and the complete proposed path. Request exact-model material and verification documents. The published manufacturer example gives concrete comparison evidence while the offered configuration retains its own acceptance decision.