Test & measurement instruments

Handheld Salinity Refractometer, 0–100 ppt NaCl & 1.000–1.070 SG

Yieryi  |  SKU: REF-S6Q8-01
$51.86 USD
Free standard shipping to supported destinations. Taxes calculated at checkout.
Package

Need help choosing? Tell us what you need to measure.

Shipping & returns

Free standard shipping is available to supported destinations. Delivery time includes order preparation and transit and varies by destination. See our shipping policy for current delivery estimates and availability.

See our return policy for return conditions. For an order-specific delivery question, contact Mcooh.

Choosing the right instrument?

Tell us your measurement range, material or target, and operating conditions.

Ask a selection question

Description

Use this handheld salinity refractometer when your salt water or brine specification is written in parts per thousand (ppt) NaCl or as specific gravity (SG). The optical eyepiece shows both scales side by side. Check the required unit before reading a sample.

Choose between NaCl and specific gravity

Eyepiece scale Range Smallest division
NaCl · right 0–100 ppt (0–10%) 1 ppt (0.1%)
Specific gravity · left 1.000–1.070 SG 0.001 SG

For a brine target written as 35 ppt, read the NaCl scale on the right. When the procedure gives a specific gravity target, use the left scale. Other dissolved substances can affect the optical reading, so use the NaCl scale for samples whose principal dissolved solid is salt. The instrument has automatic temperature compensation (ATC).

Eyepiece showing the 0–100 ppt salinity and 1.000–1.070 specific gravity scales
Read where the blue and white fields meet. Specific gravity is on the left and salinity in ppt is on the right.

Zero and read the instrument

Place distilled water on the prism and zero the instrument with the calibration screw. Wipe the prism, add the sample and close the cover plate. Read where the blue and white fields meet on the scale your procedure requires. Clean the prism between samples.

Salinity refractometer prism, cover plate, calibration screw and eyepiece
Identify the prism, cover plate, eyepiece and calibration screw before use.

Compare the three cases

Paper Box is the carton option. Plastic Case has a rigid shell, and Leather Case is the zippered option shown in the gallery. All three have the same refractometer and include a transfer pipette, calibration screwdriver, cleaning cloth and English instructions.