Multi-Probe UV Light Meter LS125 with Selected UVA, UVB or UVC Probe
Multi-Probe UV Light Meter LS125 with Selected UVA, UVB or UVC Probe - LS125 + UVC-X0 Free standard shipping to supported destinations. See the shipping policy for preparation and transit estimates.
Need help choosing? Tell us what you need to measure.
Shipping & returns
Shipping & returns
Free standard shipping is available to supported destinations. Delivery time includes order preparation and transit and varies by destination. See our shipping policy for current delivery estimates and availability.
See our return policy for return conditions. For an order-specific delivery question, contact Mcooh.
Multi-Probe UV Light Meter LS125 with Selected UVA, UVB or UVC Probe - LS125 + UVC-X0 Free standard shipping to supported destinations. See the shipping policy for preparation and transit estimates.
Choosing the right instrument?
Tell us your measurement range, material or target, and operating conditions.
Description
Description
The Linshang LS125 multi-probe UV light meter measures irradiance and accumulated energy using a detachable digital probe. Choose one of twelve probe kits for UVA, UVB, UVC or blue-light measurement. Each option supplies one LS125 main unit and one selected probe, with the probe response, aperture and measuring range matched to the light source being checked.

Choose a probe for the light source
The host recognizes the connected probe and adjusts its displayed units and measurement interface. A 1mm aperture is available for LED point sources through the UVALED-X1 option. The UVCWP-X1 option provides a waterproof probe for immersion to 1m, while the main unit stays dry.
| UVC-X0 | 230–280nm, 254nm peak. 0–200000 μW/cm². 10mm aperture. |
|---|---|
| UVCWP-X1 | 230–280nm, 254nm peak. 0–200000 μW/cm². 10mm aperture. |
| UVC-X2 | 210–250nm, 222nm peak. 0–200000 μW/cm². 10mm aperture. |
| UVCLED-X0 | 230–315nm, for 260–285nm LEDs. 0–200000 μW/cm². 10mm aperture. |
| UVB-X0 | 280–315nm, 310nm peak. 0–200000 μW/cm². 10mm aperture. |
| UVA-X0 | 315–400nm, 365nm peak. 0–2000 mW/cm². 10mm aperture. |
| UVA-X1 | 315–400nm, 365nm peak. 0–200000 μW/cm². 10mm aperture. |
| UVA-X2 | 315–365nm, 340nm peak. 0–200000 μW/cm². 10mm aperture. |
| UVALED-X0 | 340–420nm, calibrated at 395nm. 0–200000 mW/cm². 10mm aperture. |
| UVALED-X1 | 340–420nm, calibrated at 395nm. 0–200000 mW/cm². 1mm aperture. |
| UVALED-X3 | 340–420nm, calibrated at 395nm. 0–200000 μW/cm². 10mm aperture. |
| BL-X0 | 410–490nm, calibrated at 450nm. 0–2000 mW/cm². 10mm aperture. |
Irradiance, energy and response
The LCD displays real-time, maximum, minimum and average irradiance alongside duration and accumulated energy. UVA-X0, UVALED-X0, UVALED-X1 and BL-X0 sample at 2048 readings per second. The other listed probes sample at six readings per second.
For μW-range probes, accuracy is ±5μW/cm² below 50μW/cm² and ±10%H at or above that level. UVA-X0 uses ±0.5mW/cm² below 5mW/cm² and ±10%H at or above it. UVALED-X0, UVALED-X1 and BL-X0 use ±5mW/cm² below 50mW/cm² and ±10%H at or above it. H is the reference irradiance.
Host specifications
| Display | 240 x 160 dot-matrix LCD |
|---|---|
| Power | Four AAA alkaline batteries |
| Host dimensions / weight | 148 x 76 x 26mm / approx. 194g |
| Probe connection | Push-pull locking connector |
| Energy range | 0–9999999μJ/cm² for μW-range probes, or 0–9999999mJ/cm² for mW-range probes |
| Duration range | 0–99999s |
| Records | Nine saved measurement records, read on the host display |

Measurement records and handling
Use HOLD to retain a completed reading and review saved records on the instrument. Select smoothing to suit AC-powered sources when using a high-speed probe. Keep the sensing surface clean and protect it from humidity during storage. UVCLED-X0 measurements require control of strong ambient light because of the probe's wider spectral response. Position the probe with appropriate UV exposure controls in place before energizing the source.
Manufacturer documentation
Linshang LS125 user manual (PDF, English) (V3.34, 2025-01-09). Use the instructions for your selected probe.