Test & measurement instruments

Programmable DC Supply: TTL or RS-485 Control?

By Mcooh TeamPublished 7 minute read
TTL serial and RS-485 are different wired interfaces. They are not interchangeable, and a 485 variant is not automatically a Modbus device until the protocol and the register map are confirmed. RF remote control is a separate option and does not replace the wired interface. Choose the variant by the host port you actually have, then confirm the Modbus register map with us before ordering.
DPH8920 programmable DC power supply module

Quick answer: TTL serial and RS-485 are different wired interfaces. They are not interchangeable, and a 485 variant is not automatically a Modbus device until the protocol and the register map are confirmed. RF remote control is a separate option and does not replace the wired interface. Choose the variant by the host port you actually have, then confirm the Modbus register map with us before ordering.

Separate three decisions: voltage class, wired interface, and RF

The DPH8909 and DPH8920 families are DC-input programmable buck modules. Both list a 20–110 V DC input and a 0–96 V adjustable output. They are not designed for direct connection to AC mains; they need an appropriate DC source in the stated input range. Actual output depends on the input source, the load and cooling. Do not assume that the maximum voltage, current and power are available at the same time.

The current rating is the first split: DPH8909 is listed at 0–9.6 A and 0–921.6 W; DPH8920 is listed at 0–20 A and 0–1920 W. After that, each family has four communication packages: TTL only, TTL with RF, 485 only, and 485 with RF.

Texas Instruments’ RS-485 Basics (SLLA545) describes RS-485 as a differential physical layer, not an application protocol. Many upper-layer protocols can run on that physical layer. TTL/UART signalling is a different electrical interface and needs a transceiver to reach an RS-485 A/B pair. That is why a TTL SKU and a 485 SKU are not drop-in substitutes for each other.

Block diagram showing a host connecting either through a TTL path or through a separate RS-485 path, with RF drawn as an optional extra
TTL and RS-485 are separate wired interfaces. RF is an option on top, not a substitute. Schematic, not measured; pinouts are not shown.

The four packages in each family

Variant pattern Wired interface RF remote DPH8909 SKU / price (USD) DPH8920 SKU / price (USD) Listed control accessories
Base TTL serial No INS06-PS3317, 162.52 INS06-PS3321, 191.28 Module and USB-to-TTL cable
RF TTL serial Yes INS06-PS3315, 213.36 INS06-PS3319, 233.42 Module, remote, remote charging cable, and USB-to-TTL cable
485 485 interface No INS06-PS3318, 177.84 INS06-PS3322, 191.21 Module and USB-to-485 module
485RF 485 interface Yes INS06-PS3316, 236.11 INS06-PS3320, 264.54 Module, remote, remote charging cable, and USB-to-485 module

RF features apply only to variant names that contain RF. The product pages state that PC control supports SIMPLE and MODBUS communication through the applicable interface, and they ask you to confirm software compatibility, drivers and protocol details for the selected variant. This article’s confirmation point is the Modbus register map: a 485 interface does not, by itself, define the register addresses or the function codes your host will use.

The listing does not include pinouts, termination values, or a guaranteed communication distance. The RF variants list 2.4 G, a 10 m communication distance under the stated accessible condition, channels 01–30 and addresses 00–99; treat those figures as the listing, not as a site survey.

Shared electrical limits, different current class

Both families list a voltage resolution of 10 mV. Current resolution is 1 mA on DPH8909 and 10 mA on DPH8920. Display accuracy follows the same split (10 mV and 1 mA, or 10 mV and 10 mA). Shared series figures on the product pages include: power effect CV <0.5% + 10 mV and CC <1% + 5 mA; load effect in the same terms; output ripple <50 mVpp at 40 V input, 12 V output, 7 A; typical efficiency 92% at 40 V input, 30 V output, 7 A; output error voltage ±2‰ + 1 character and current ±5‰ + 2 characters; response time <50 ms; 10 parameter groups (M0–M9); PT over-temperature protection; onboard radiator and fan cooling; minimum voltage difference 0.5 V. Module dimensions are listed at 150 × 120 × 59 mm.

Those shared figures do not change the interface choice. Size the current class first, then pick TTL or 485 from the host, then add RF only if you need the wireless remote.

Four-cell matrix of TTL or RS-485 along one axis and RF or no RF along the other, for the DPH8909 and DPH8920 families
Pick current class, then the wired interface, then whether RF is required. Schematic, not measured.

Worked example: a PLC with an RS-485 port

A machine controller exposes an RS-485 port and the load needs about 15 A at a voltage inside the 0–96 V window. The current class points to DPH8920, not DPH8909 (9.6 A). The wired interface points to a 485 variant (INS06-PS3322 or, if a remote handset is also required, INS06-PS3320). A TTL SKU will not plug into that RS-485 port. Before the host software is written, the Modbus register map still has to be confirmed, because the presence of a 485 interface does not publish the map.

The DC input source must stay inside 20–110 V and stay high enough for the intended output, with cooling sized for the load. The module is a buck converter: it is not an AC-mains instrument.

A short selection sequence follows the same split:

  1. Current class — 0–9.6 A / 0–921.6 W (DPH8909) or 0–20 A / 0–1920 W (DPH8920), sized with input and cooling, not by multiplying the three maxima.
  2. Wired interface — TTL serial or 485, according to the host port. They are not interchangeable.
  3. RF option — only if a listed RF variant is required; it does not replace the wired port.
  4. Protocol — the pages mention SIMPLE and MODBUS on the applicable interface. Confirm the Modbus register map as already stated, and treat drivers and software compatibility as part of the same order check.
  5. Package — each SKU lists its own cable or USB-to-485 module, and RF variants add the remote and its charging cable.

Ten parameter groups (M0–M9) can store voltage and current limits on the module itself. That local memory is not a communication protocol and does not remove the need for the correct wired interface. Current modules are in the programmable DC power supply module collection.

Worked example: a bench host with a USB-TTL port

A test bench already has a USB-to-TTL adapter in daily use, and the load is a 6 A, 48 V fixture well inside the DPH8909 current class (0–9.6 A, 0–921.6 W). The wired interface points to a TTL variant: INS06-PS3317 if no remote is needed, or INS06-PS3315 if the listed RF handset is also required. A 485 SKU would be the wrong electrical interface for that host. The DC input still has to sit in 20–110 V and stay at least 0.5 V above the 48 V output, with cooling sized for the load. Local memory groups M0–M9 can store the 48 V / 6 A limit on the module; they do not replace the wired port, and they do not publish a Modbus map.

Mistakes that show up on real jobs

Plugging a TTL module into an RS-485 port, or the reverse, and expecting a different cable to fix it. The interfaces are different electrical layers. Texas Instruments’ SLLA545 describes RS-485 as a differential physical layer, not an application protocol.

Treating a 485 variant as a finished Modbus device. The product pages say PC control supports SIMPLE and MODBUS on the applicable interface. The register map is the confirmation already stated.

Buying an RF variant in place of the wired interface. RF is listed only on named RF SKUs, and it does not replace TTL or 485.

Multiplying 96 V by 20 A and assuming 1920 W is always available on a DPH8920, or 96 V by 9.6 A as 921.6 W on a DPH8909. The listings give those maxima and also state that actual output depends on the input source, the load and cooling.

Connecting the module to AC mains. Both families are DC-input buck modules with a 20–110 V DC input. They need a separate DC source in that range, with the 0.5 V minimum difference above the intended output.

Omitting cooling or stacking modules so that the fan and radiator cannot do the work. Over-temperature protection is listed; it is not a substitute for sizing the installation.

FAQ

Can I use a TTL module on an RS-485 bus with a different cable?

No. The electrical interfaces are different. Choose the 485 variant, or add only a conversion arrangement that you have specified and tested yourself. This listing does not include such a converter as a standard TTL accessory.

Does a 485 variant guarantee Modbus?

No. The pages say PC control supports SIMPLE and MODBUS on the applicable interface. The register map still has to be confirmed for the selected variant.

Does the RF remote replace the wired port?

No. RF is an option on named RF variants. The wired TTL or 485 interface is chosen separately.

Can I draw 96 V and 20 A and 1920 W at the same time from a DPH8920?

The listing gives those maxima, and it also states that actual output depends on the input source, load and cooling. Do not assume the three maxima are available together.

Sources

Specifications and prices are from the linked product pages and manufacturer documents and may change. Follow the supplied manual and the procedures that apply to your work.

More from Power Supply Guides

Need help choosing an instrument?

Tell us what you need to measure, the range and the application. You can also browse our measurement guides.