Description

The Linshang LS163 and LS163A window tint meters compare the optical properties of solar film, film-on-glass, compatible automotive side glass and other uniform transparent materials. Choose LS163 for three-parameter checks, or the seven-parameter LS163A for additional infrared and solar-energy readings. Both use a parallel light path and a 6mm-wide measuring slot.

LS163A optical parameter diagram listing UV visible IR Full IR TSER and SHGC
LS163A optical parameter diagram listing UV visible IR Full IR TSER and SHGC

Choose LS163 or the seven-parameter LS163A

LS163 measures visible light transmission, UV rejection or transmission, and infrared rejection or transmission at 940nm. Its three display layouts provide VLT with UVR/IRR, VLT with UVT/IRT, or VLT alone.

LS163A provides seven available optical parameters. Alongside visible light and UV, it provides infrared readings at 940nm and 1400nm, Full IR, total solar energy rejection (TSER) and solar heat gain coefficient (SHGC). Select the displayed parameter combination to suit the comparison. Record the chosen channel or solar-energy parameter with the result so measurements can be compared consistently.

Measurement LS163 LS163A
Visible light 380–760nm, CIE photopic response 380–760nm, CIE photopic response
UV source 365nm 365nm
Infrared channels 940nm 940nm, 1400nm and Full IR
TSER range — 0.0–100%
SHGC range — 0.000–1.000

Fit the sample and establish the reference

The 6 x 58mm slot supports loose film and suitably sized glass samples. Check that the specimen fits the opening and reaches the optical measuring position. Use a clean, uniformly transparent test area and consistent placement when comparing samples.

Switch on with the slot empty and allow self-calibration to finish before inserting a specimen. The stated ±2% accuracy applies to colorless, transparent materials. Strong ambient light, an obstructed optical path or dust can prevent startup calibration. Keep the opening clear and follow the operating procedure when establishing a new reference.

LS163A selectable display combinations for transmission rejection TSER and SHGC
LS163A selectable display combinations for transmission rejection TSER and SHGC

Select a display and hold the reading

The backlit display rotates with the measuring orientation. On LS163A, a short button press with no sample in place switches the display interface. During measurement, a short press holds the result for viewing after the sample is removed. LS163 provides its three display layouts through the settings menu and also supports data hold.

Shared instrument specifications

Test slot 6mm wide x 58mm deep
Resolution 0.1%
Accuracy ±2% for colorless, transparent materials
Display 128 x 128 dot-matrix LCD with backlight and automatic rotation
Power supply 4 AAA alkaline batteries
Dimensions 146 x 73 x 24mm
Instrument weight Approximately 187g including batteries