UV LED Energy Meter LS128 with Irradiance and Temperature Curves
UV LED Energy Meter LS128 with Irradiance and Temperature Curves Free standard shipping to supported destinations. See the shipping policy for preparation and transit estimates.
Need help choosing? Tell us what you need to measure.
Shipping & returns
Shipping & returns
Free standard shipping is available to supported destinations. Delivery time includes order preparation and transit and varies by destination. See our shipping policy for current delivery estimates and availability.
See our return policy for return conditions. For an order-specific delivery question, contact Mcooh.
UV LED Energy Meter LS128 with Irradiance and Temperature Curves Free standard shipping to supported destinations. See the shipping policy for preparation and transit estimates.
Choosing the right instrument?
Tell us your measurement range, material or target, and operating conditions.
Description
Description
The Linshang LS128 UV LED energy meter measures irradiance, accumulated energy and temperature during LED curing. Its 340–420 nm spectral response is calibrated at 395 nm and supports checks on UV LED sources such as 365, 385, 395 and 405 nm lamps.

Capture the UV LED exposure profile
The current LS128 measuring range reaches 200000 mW/cm² with 1 mW/cm² resolution. A 32-minute run records the exposure while the display provides real-time and maximum values. Irradiance and temperature curves can be reviewed after measurement to examine changes through the process.
Use manual mode to start and stop a test directly. Automatic mode waits for the selected irradiance threshold, records while the trigger condition is met and stops when the signal falls below the threshold or the recording period is reached. This helps align recording with the lamp exposure during conveyor checks.
LS128 spectral and recording data
| Spectral response | 340–420 nm; calibrated at 395 nm UV LED |
|---|---|
| Irradiance range | 0–200000 mW/cm² |
| Irradiance resolution | 1 mW/cm² |
| Accuracy for H ≥ 50 mW/cm² | ±10%H, where H is the reference value |
| Energy range | 0–999999 mJ/cm² |
| Sampling speed | 2048 samples/second |
| Recording period | 32 minutes |
| Temperature measurement | -55°C to +125°C |
| Power supply | 2 AAA alkaline batteries |
| Display | 240 x 160 dot-matrix LCD |
| Dimensions | 120 mm diameter x 13 mm thick |

Read curves and export process records
Connect the USB interface to Windows PC software in the stopped state to read the recorded data. The software supports curve review, spreadsheet export and report generation. The latest measurement remains stored after the meter is switched off.
Record the LED wavelength, measuring position, trigger mode and relevant smoothing settings with the results. Using the same optical conditions makes repeat measurements easier to compare across curing runs.
Orient the detector toward the source
The sensor is on the rear face of the disc. Keep the sensing area clean and place it consistently during each test. The separate temperature channel covers -55°C to +125°C; use the instrument within its operating conditions and store it in a dry environment between measurements.